EndoActivator® System
Technique Card
Activate Your Endodontics
Technique Sequence
Sequence for cleaning root canal systems. For other applications, see Directions For Use.
- Prepare canal to produce a fully tapered shape
- Fill pulp chamber with NaOCl, EDTA, or other final rinse solution
- Select the Activator tip that manually fits loosely within 2 mm of working length
- Place the barrier sleeve over the Driver to protect the entire hand piece
- Attach the Activator tip over the barrier-protected Driver
- Place the attached Activator tip into the prepared root canal
- Depress the on/of switch to activate solution
(Note: Switch defaults to high speed, as indicated for cleaning procedures) - Use a pumping action to move the Activator in short, 2-3 mm vertical strokes
- Hydrodynamically agitate the intracanal solution for 30-60 seconds
- Irrigate, then use intracanal suction to eliminate loose debris
- Repeat the above steps for each intracanal irrigant used
Product Features
EndoActivator Tips
- Snap-on/snap-off design
- Strong, flexible, medical-grade polymer composition
- Color-coded by size
- 22 mm lengths with depth gauge rings at 18, 19 and 20 cpm
Hand piece/Driver
- Contra-angled design for easy access to posterior teeth
- 3-speed sonic motor provides options of low, medium and high
- Ergonomic, cordless, and batteryoperated hand piece

See Directions For Use
For Dental Use Only